Ady Water menawarkan conductivity/konduktiviti meter dari Brand /Merek Horiba yaitu tipe HE-300C. Alat conductivity meter ini cocok untuk pengukuran terus-menerus air murni atau air pendidih, yang menjadi keunggulannya adalah kompensasi temperatur yang flexibel dengan rentang lebar.

Rentang Pengukuran Cell constant /cm 0.01 0.1 1
Electrical conductivity μS/cm 0.000 to 2.000; 0.00 to 10.00 0.000 to 2.000; 0.00 to 20.00; 0.0 to 100.0 0.0 to 200.0; 0 to 1,000
mS/m 0.0000 to 0.2000; 0.000 to 1.000 0.0000 to 0.2000; 0.000 to 2.000; 0.00 to 10.00 0.00 to 20.00; 0.0 to 100.0
TDS conversion mg/L 0.00 to 2.00; 0.0 to 10.0 0.0 to 20.0; 0 to 100 0 to 200; 0 to 1,0000
Electrical conductivity measurement and TDS conversion measurement cannot be selected at same time.
Temperature ˚C 0 to 100; Resolution: 0.01˚C
Pengulangan Electrical conductivity: ±0.5% full-scale or less
TDS conversion: ±1.5% full-scale or less; Temperature: ±0.1˚C (for equivalent input)
Linearitas Electrical conductivity: ±0.5% full-scale or less; TDS conversion: ±1.5% full-scale or less; Temperature: ±0.5˚C (for equivalent input)
Transmission output 4 to 20 mA DC   Input/output insulated type; Maximum load resistance 600 Ω
Contact input One point
Contact format: Non-voltage a contact;
Contact function: Transmission output is held for closed contact input.
Rentang Kompensasi Temperature   0 to 100˚C
(but for 0˚C or less, 100˚C or more, extend and perform calculation)
Ambient temperature -20 to 60˚C
Temperature compensation element Pt 1000 (0˚C)
Calibration function Electrical conductivity: Based on input of cell constant correction coefficient (parameter input); TDS conversion: Conversion based on specified coefficient (0.30 to 1.00); Temperature: One point calibration compared with reference temperature gauge
Self-diagnosis function Sensor diagnosis error, Transmitter malfunction
Power supply 24 V DC (operational voltage range: 21 V to 32 V DC); Power consumption 0.6 W or less
Construction Outdoor installation type: IP65 protection level; Mounting method: 50A pole or wall mounted; Case: Aluminum alloy   Mount fitting/hood: SUS304
Compatible sensor ESH series electrical conductivity sensor (cell constant 0.01/cm, 0.1/cm, 1.0/cm)
Weight Approx. 4 kg
Equipment protected CE marking, FCC rules